11
Chapter 1: General Information
Specifications
The specifications are the performance standards against which the product is
tested.
Specifications
Parameter
300 Mb/s
mode
600 Mb/s
mode
Notes
Minimum master to
master clock time
3.33 ns
1.67 ns
tWidth (minimum)
1 ns
1 ns
Specified at probe tip. Eye width as measured
by Eye Finder may be less.
Specifications verified under the following test conditions:
Parameter
300 Mb/s
mode
600 Mb/s
mode
Notes
Vh
1.125 V
250 mV pp
Vl
0.875 V
vThreshold
1 V
rise/fall times
150-180 ps
Probe
Agilent E5382A
tWidth
(Data Eye)
vHeight
Individual
Data Channel
vThreshold
0V
Sampling
Position
(Eye Finder
Blue Bar)