99
Chapter 5: Troubleshooting
To test the cables (16900-series Mainframe)
This test allows you to functionally verify the logic analyzer cable and an Agilent
E5379A probe.
Equipment Required
1
Connect the logic analyzer to the stimulus board.
a
Connect an Agilent E5379A 100-pin differential probe to the logic
analyzer cable (also called “Pod”) to be tested. Start with Pod 1.
b
Connect the E5379A probe to connector “Pod 4” on the stimulus board.
c
Connect the stimulus board power supply output to the stimulus board
power supply connector J82.
d
Plug in the stimulus power supply to line power. The green LED DS1
should illuminate showing that the stimulus board is active.
Equipment
Critical Specification
Recommended Part
Stimulus Board
No Substitute
16760-60001
Differential Probe
No Substitute
E5379A