38
Chapter 3: Testing Logic Analyzer Performance
Testing the Module Using a 16900-Series Mainframe
Do a self-test on the 16900-series logic analysis system
a
When the logic analysis system has finished booting, the Waveform
window appears. Select
Help
→
Self-Test...
from the main menu. The
Analysis System Self Tests window will appear.
b
In the Select Suite(s) list, select <all>. This will cause <all> to be
selected in the Select Test(s) list.
c
Select
Start
. This will perform a complete system self-test.
The progress of the self tests is displayed in the Progress & Statistics area of the
window.
d
When the self-tests are complete, check the Results window to ensure
that the Result Summary says that all tests passed. If all tests did not
pass, refer to “Troubleshooting” on page 89.
e
Select the
Close
button to close the Analysis System Self Tests window.
f
If all module self-tests pass, then record “PASS” in the “Logic Analysis
System Self-Tests” section of the Performance Test Record (page 60).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com