39
Chapter 3: Testing Logic Analyzer Performance
Testing the Module Using a 16900-Series Mainframe
Configure the 16900-Series Logic Analysis System
1
Exit the logic analysis application (from the main menu, choose
File
→
Exit
)
and then restart the application. This puts the logic analysis system into its
initial state.
2
Disable all logic analyzers other than the analyzer under test.
a
Select the
Overview
tab at the bottom of the main window.
b
Click on each unused logic analyzer and select disable. Only the logic
analyzer to be tested should remain enabled.
3
Set up the bus and signals.
a
From the Logic Analysis System main menu, select
Setup
→
My
16950A
→
Bus/Signal...
.
b
In the Analyzer Setup window, choose the
Threshold
button for Pod 1.
The Threshold Settings window will appear.