106
Chapter 5: Troubleshooting
2
Set up the stimulus board
a
Configure the oscillator select switch S1 according to the following
settings:
•
S1 0 (Off)
•
S2 1 (On)
•
S3 0 (Off)
•
Int
b
Configure the data mode switch S4 according to the following settings:
•
Even
•
Count
c
Press the Resynch VCO button, then the Counter RST (Counter Reset)
button.
3
Set up the logic analyzer
a
Open the Session Manager window and select “Start Session.”
b
In the Logic Analysis System window, select the module icon, then
select Listing. A Listing window appears. You will use this window later.
c
In the Logic Analysis System window, select the module icon again,
then select Setup and Trigger. A Setup and Trigger window appears.
4
Set up the Sampling tab
a
In the logic analyzer Setup and Trigger window, ensure that the
Sampling tab is selected.
b
Under the Sampling tab, select State Mode.
c
In the upper right area of the window, select the Timing Zoom... button
d
Switch Timing Zoom off, and close the Timing Zoom window.
e
In the Clock Setup area of the Sampling tab, ensure that the Mode is set
to Master only.
f
In the Clock Setup area of the Sampling tab, ensure that the clock edge