Chapter 5: Troubleshooting
analyzer to be tested should remain enabled.
5
From the Logic Analysis System main menu, select
Setup
→
My
16950A
→
Timing/State (Sampling).....
a
Select
State Mode
.
b
Clear the Timing Zoom check box to turn Timing Zoom off.
c
Set the Trigger Position to
100% Poststore
.