Chapter 3: Testing Logic Analyzer Performance
Testing the module using a 16700-series Mainframe
c
Select the Acquisition Depth field, then select 128K.
d
Select the Trigger Position field, then select Start.
e
Select the Count field, then select Off.
3
Configure the Format settings.
a
In the logic analyzer Setup and Trigger window, select the Format tab.
b
Under the Format tab, select Pod Assignment.
c
In the Pod Assignment window, use the mouse to drag the pods from
the Analyzer 2 column to the Analyzer 1 column.
d
Select Close to close the Pod Assignment window.
e
Under the Format tab, use the scroll bar to scroll to the fields for Pod
being tested (Pod 1).
f
Select the field showing the channel assignment for the pod being