Chapter 3: Testing Logic Analyzer Performance
Test Pod 2 in 300 Mb/s Mode
12
Select the
OK
button to close the “Threshold Settings: Clock for Pod 2”
window.
13
Select the
OK
button to close the “Threshold Settings - Clocks” window.
14
Select the Sampling tab (at the top of the window). In the State Options
area, set clock Clk1 to
Don’t Care.
15
Set Clk2 to
Both Edges
.
16
Adjust the sampling positions using Eye Finder. Be sure to expand “My
Bus 1” and use the recommended starting position noted on page 44.
Realign any stray channels if necessary. See page 45.
17
Determine pass or fail (1 of 2 tests). See page 46.
18
Select
OK
to close the “Analyzer Setup” window.
19
Switch to the Listing window by selecting the
Listing
tab at the bottom of
the main logic analyzer window.
20
Select the Run Repetitive icon
.
21
Determine pass or fail (2 of 2 tests). See page 48.
Test the complement of the bits (Pod 2, 300 Mb/s mode)
1
Test the complement of the bits. See page 49.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com