Mains supply
Topics:
•
•
•
Radiographic table
All the equipments of any X-ray installation must be connected to the
electrical mains by means of a contact or another multi-polar sectioning
device that must be installed by the customer. The room of the installation
must be in compliance with the IEC/CEI or UL/CSA prescriptions in force
(where requested). The relative national regulations must be applied in all
countries. Nonetheless, compliance with the above- mentioned standards is
strongly recommended, when they do not contradict the national regulations
in force, to ensure safety for the operator, the patient and third parties.
Table 61: Radiographic table
Rated voltage
230 Vac
Voltage change
220/240 V
Allowed fluctuation of mains voltage ± 10%
Mains frequency
50/60 Hz
Electrical current absorption
15 A
Thermal dissipation
500 W
Type of protection
Class I
Degree of protection
Type B
Degree of casing protection
IPX0
Electromagnetic compatibility
Equipment in compliance with EN
60601-1-2 standards
Normally the power supply of the remote-controlled table is picked up by a
16A, delayed 400Vac nominal bipolar sectioning automatic switch, placed in
the mains supply panel board of the exam room, with intervention differential
protection not higher than 30mA in compliance with IEC/CEI standards. The
ground terminal of the equipment must be connected in a visible manner to
the earth equipotential node of the diagnostic system by means of a protection
conductor, with the minimum section of 10 mm
2
. The power cable from the
mains supply panel board to the equipment electrical cabinet must have a
minimum section of 2.5 mm
2
. All the power connections and earth cables
must be in compliance with the national rules related or equivalent.
DR 800 | Product Information |
307
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...