Disinfecting safety directions
WARNING:
Using a disinfectant that can form an explosive or flammable
gas mixtures is hazard to life and health because of explosion
risk. Switch the equipment off before disinfecting. Allow the gas
mixture to evaporate before switching the x-ray system back on.
CAUTION:
Using unsuitable disinfectants can cause discoloration and
damage of the surface of the equipment. If a functional
degradation or malfunctioning of the product is noticed due to
disinfection, contact the medical device manufacturer.
To disinfect the device:
• Do not use any corrosive, soluble or gaseous disinfectants.
• Consult the manufacturer’s Material Safety Data Sheets (MSDS) and
recommendations on the product label for additional information prior to
use.
• Use of spray disinfection can cause malfunctions due to ingress of the
disinfectant into the equipment. Disinfect all parts of the unit, including
the accessories and connection cables by just wiping them. Switch off the
system and cover the cooled system carefully before performing a room
disinfection using nebuliser.
DR 800 | Cleaning and Disinfecting |
87
0392C EN 20210309 1049