FLFS overlay
The FLFS overlay consists of a stitching grid and a mounting unit. The
stitching grid must be attached to the radiographic table using the mounting
unit.
The partial images of the FLFS examination are automatically stitched to
create the DR Full Leg Full Spine image. Stitching is applied based on a grid of
markers inside the stitching grid and a correction is applied based on the
alignment of the anatomical information in the image.
1
2
1.
Mounting unit
2.
Stitching grid
Figure 22: FLFS overlay
40
| DR 800 | Introduction
0392C EN 20210309 1049