Performing offset calibration
Figure 151: Offset Calibration button
The
Offset Calibration
button reminds the user to manually trigger the offset
calibration in case the system cannot perform the offset calibration
automatically.
It is required to manually trigger the offset calibration, in order to prevent
image artifacts caused by delayed offset calibration.
1.
Press the offset calibration button in the DR Detector switch.
The X-ray modality controls are displayed on the software console.
2.
Press the offset calibration button.
The offset calibration is performed.
DR 800 | Software Console |
195
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...