Figure 107: Generator controls for fluoroscopy and for roadmapping
WARNING:
Higher frame rates (FPS) for DSA lead to shorter
examination times until the heat limit is reached. Lower
frame rates are recommended where applicable, especially
in combination with thicker or more absorbing body parts
3.
Move the X-ray system to the right position.
a) Check if a correct automatic position is selected.
Figure 108: Positioning controls on the software console
b) Move the X-ray system to the selected automatic position.
The actual and target position parameters are displayed on the
software console. When the target position is reached, the movement
stops.
c) Adjust the position using the position controls.
4.
Position the patient.
Use patient fixations available to avoid patient movement during
interventional procedures.
The patient position can be verified using the collimator camera.
5.
On the collimator, switch on the light localizer. Apply collimation.
6.
Acquire a set of DSA sequences, roadmapping examinations, fluo
sequences and static images.
Any number of DSA sequences, roadmapping examinations, fluo
sequences or static images can be acquired, in any order.
Table 10: Supported workflows
Image
type
Settings Step 1: activate
Step 2: start
exposure
Result
Road-
map-
ping
After the final expo-
sure, finish the
workflow using the
same button:
fluoroscopy
pedal
Roadmapping
mask:
One or more
roadmapping
sequences
132
| DR 800 | Basic Workflow
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...