1.
Patient orientation
2.
Mounting unit
3.
Tabeltop guides
4.
FLFS overlay
Figure 126: FLFS overlay on radiographic table (head left, head right)
3.
Use the positioning console to move the X-ray system to the automatic
position.
The X-Ray system moves to the default position for preparing the DR Full
Leg Full Spine examination.
If the radiographic table is tilted to vertical position, assure that the FLFS
overlay remains flat against the tabletop.
CAUTION:
After doing vertical FLFS exposures the grid can swing away
from the table and disturb the patient. Set the inclination of
the table to 89 instead of 90 degrees.