Tests of Re-
sistance to
Disruption
IEC
60601-1-2:2014
Test Level
Level of
Agree-
ment
Electromagnetic Environ-
ment
Conducted
RF
IEC
61000-4-6
3 V
150 kHz to 80 MHz
3 V
150 kHz
to 80
MHz
Portable and mobile RF com-
munications equipment
should be used no closer to
any part of the device, in-
cluding cables, than the rec-
ommended separation dis-
tance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d
= 1.2
150 kHz to 80
MHz
d
= 1.2
80 MHz to 800
MHz
d
= 2.3
800 MHz to 2.5
GHz
where ’P’ is the maximum
output power rating of the
transmitter in watts (W) ac-
cording to the transmitter
manufacturer and ’d’ is the
recommended separation dis-
tance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz
to 2.5
GHz
DR 800 | Product Information |
321
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...