Error messages on the ceiling suspension
Table 51: Error messages are displayed on the display of the control panel of
the X-ray tube of the ceiling suspension
Error
message
Possible cause
Possible solution
FAULT EE-
PROM 567
Read Only Memory
(EEProm) of the PCB
567 of the ceiling
suspension unit main
CPU has stopped
working.
Stop use of the system and call technical
service. This may require a PCB replace-
ment.
FAULT EE-
PROM 628
Read Only Memory
(EEProm) of the PCB
628 of the ceiling
suspension unit con-
trolbox has stopped
working.
Stop use of the system and call technical
service. This may require a PCB replace-
ment.
SHORT
CIRC. 567
Brake short circuit to
ceiling suspension.
Stop use of the system and call technical
service. A hardware device has failed.
The lateral SID is un-
known.
The ceiling suspension must be set into
one of three programmed positions to
display the lateral SID (with digital
switches).
If the SID does not display at one of the
positions, a mechanical switch may be
damaged, its connection wire discon-
nected, or it is out of alignment. The sys-
tem will only allow manual controls.
Contact technical service for support.
274
| DR 800 | Problem solving
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...