1.
Relesae the tilting brake, turning the knob of the tilting brake
counterclockwise.
2.
Adjust the tilting angle.
3.
Lock the bucky into position, turning the knob clockwise.
To rotate the bucky:
1
2
3
1.
Detector tray lock
2.
Detector tray
3.
Rotation lock lever
Figure 28: Rotation controls
1.
Make sure the detector tray is locked, to protect the DR detector from
falling.
2.
Hold the bucky with one hand and with the other hand release the lever
that locks the rotation.
3.
Rotate the bucky to the right orientation and release the lever.
4.
Continue the rotation until the lock closes.
WARNING:
Verify that the bucky tray has been blocked firmly into position
when making rotations! If the device has not been locked down,
it may be ejected from the rotation movement and fall to the
ground causing damage to it or injury to the patient and/or the
operators!
WARNING:
Never extract bucky tray when the loading position is pointed
towards the ground! It can easily fall to the ground causing
damage to it or injury to the patient and/or the operators!
WARNING:
Always keep the rotation of the bucky under control. Do not
rotate the bucky in an uncontrolled manner or allow it to
forcefully strike the mechanical stops. This could damage the
components or cause premature failures.
46
| DR 800 | Introduction
0392C EN 20210309 1049
Содержание 6010/200
Страница 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Страница 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Страница 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Страница 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Страница 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...