8250A Optical Power Meter Operation Manual
3.5 Wavelength Preset
3-9
3.5
Wavelength Preset
Frequently used wavelengths can be set easily by registering the four wavelengths.
The preset wavelength cannot be set and called by using the remote command.
3.5.1
Setting the Preset Wavelength
The four wavelengths from No.0 to No.3 can be set. The wavelength setting range is from 0 nm to 9999
nm.
An example, in which No.1 is set to 405 nm, is shown below. See Figure 3-5.
To cancel the setting, press
[CE]
.
Figure 3-5 Setting the Preset Wavelength
Key operation
Display
1.
[MENU]
010
SM.CNT
The first item of the menu is displayed.
2.
[PREV]
5 times
0. 0780
P.SET
The current setting of the No.0 preset wavelength
is displayed.
The least significant digit blinks. (Cursor)
3.
[
]
0. 0780
P.SET
The most significant digit blinks. (Cursor moves)
4.
[
]
1. 0780
P.SET
The current setting of No.1 is displayed.
5.
[
]
1. 0780
P.SET
The least significant digit blinks. (Cursor moves)
6.
[
]
[
]
1. 0785
P.SET
The cursor under the wavelength digit moves.
7.
[
]
[
]
1. 0405
P.SET
Adjust the value to 405.
8.
[ENTER]
Normal display
Setting complete.
Normal display
Determining
the setting
Canceling
the setting
Preset wavelength
Wavelength setting range
0 to 9999
Initial value when shipping
All wavelengths are set to 780
5 times
7 times
Moving the item or cursor or changing the value
Wavelength number
Preset wavelength