8250A Optical Power Meter Operation Manual
6.1.1 How to Connect Instruments
6-2
6.1.1
How to Connect Instruments
Figure 6-1 shows the connection diagram required for the performance test of this instrument.
Figure 6-1 Connection Diagram for the Performance Test
6.1.2
How to Test
Perform the performance test in an area free from dust, vibration, and noise, and under the following envi-
ronmental conditions:
Temperature:
23
C
5
C
Relative humidity: 70% or less
Warm-up time:
5 minutes or more
The performance of this instrument (8250A) can be checked by performing a current measurement test.
This instrument allows the different internal measurement path (sensor series path) according to the sensor
series (823* or 821* series) connected to this instrument. Because the specifications are different between
each sensor series, switch the path and perform the test.
The path can be switched by replacing the TRIAX-12pin conversion adapter A (for the 823* series) with
TRIAX-12pin conversion adapter B (for the 821* series).
Procedure
1. Connect the 823* series sensor to this instrument and turn on the power.
2. By pressing the
[MENU]
key and
[PREV]
,
[PREV]
,
[PREV]
, and
[
]
, CAL
ON blinks. Press the
[ENTER]
key to set the CAL mode.
(Even if the instrument enters the CAL mode, the calibration data remains
unchanged.)
The instrument enters the current measurement mode.
Input open
TRIAX-12pin
conversion
adapter A or B
Connection for zero
measurement
Connection for the full-scale
measurement
TRIAX-12pin
conversion
adapter A or B