8250A Optical Power Meter Operation Manual
A.2 Error Messages
A-4
Table A-4 Causes of the Generated Calibration Errors
Cause
Command
Display
Status
Remarks
1
The I/V offset calibration is not con-
vergent.
XIVC
Err 10
ERR Bit7 ON
2
The difference between the VIG gen-
eration value, which is input when the
zero or full scale calibration is per-
formed, and the measured value
exceeds 50% of the full scale.
XPC
Err 10
ERR Bit7 ON
The value measured when calibrat-
ing exceeds a value allowed in the
measurement range.
3
The full scale calibration is per-
formed without zero calibration.
XPC
Err 15
ERR Bit13 ON
Command execution
error
4
The VIG generation value is not
within the input range.
XPC
Err 15
ERR Bit13 ON
Command execution
error
5
Calibration is complete.
I/V offset calibration
XIVC
DONE
DSR Bit2 ON
ZERO/FULL calibration
XPC
Writing the calibration data
XWR
Writing the sensor calibration data XSWR