7351 Series Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-29
System
SRQ
S1
Prohibition of SRQ transmitting
S0
Permission of SRQ transmitting
S?
Reply: S0 or S1
-
Status
*CLS
Clears each status byte.
CS
Status
*STB?
Reads the status byte register
Reply: ddd
*SREn
Sets the service request enable register (SRER).
n: 0 to 255 (However, bit6 cannot be set.)
-
*SRE?
Reply: ddd
-
*ESR?
Reads the standard event status register (SESR).
Reply: ddd
-
*ESEn
Sets the standard event status enable register (SESER).
n:0 to 255
-
*ESE?
Reply: ddd
-
MSR?
Reads the measurement event register (MER).
Reply: ddddd
-
MSEn
Sets the measurement event enable register (MEER).
n:0 to 65535
-
MSE?
Reply: ddddd
-
QSR?
Reads the questionable event register (QER).
Reply: ddddd
-
QSEn
Sets the questionable event enable register (QEER).
n:0 to 65535
-
QSE?
Reply: ddddd
-
OSR?
Reads the operation event register (OER).
Reply: ddddd
-
OSEn
Sets the operation event enable register (OEER).
n:0 to 65535
-
OSE?
Reply: ddddd
-
*PSCn
n:-32767 to +32767
If "n" is set to any value except for zero, SRER and SESER
are cleared when the power is turned on.
If "n" is set to zero, SRER and SESER are not cleared when
the power is turned on.
-
*PSC?
Reply: 0 or 1 (When any value except for zero is set.)
-
Option
*OPT?
Reads the option information.
Reply: Option name (RS-232, COMPARATOR) or 0
-
Operation complete
*OPC
Sets the operation complete bit (bit0) of the Standard Event
Status Register after all operations are complete.
-
*OPC?
Reply: 1 (After all operations are complete.)
-
*WAI
Waits for the completion of all operations. (GPIB only)
-
Item
Command
Description
Initial
value
R6451