![ADCMT 7351A Скачать руководство пользователя страница 179](http://html1.mh-extra.com/html/adcmt/7351a/7351a_operation-manual_2848271179.webp)
7351 Series Digital Multimeter Operation Manual
8.11 Error Message List
8-23
19
Err216
1/500 ATT measurement test error in analog operation
[ATT 1/500 Check]
Repair
20
Err500
Calibration data error
[CAL data]
Solution3
21
Err501
ZERO calibration is not performed.
[before ZERO]
Solution4
22
Err-102
Command syntax error
[Cmd Syntax]
Solution5
23
Err-113
The command cannot be used.
[Undefined header]
Solution5
24
Err-121
Unrecognized character as the numerical value
[Invalid character in number]
Solution5
25
Err-141
Invalid character string parameter is entered.
[Invalid character data]
Solution5
26
Err-151
Error as the character string data (Quotation mark error)
[Invalid string data]
Solution5
27
Err-200
Execution error (Command which cannot be executed now)
[Execution error]
Solution6
28
Err-213
The INITiate command is ignored because the measurement
is performing.
[Init ignored]
Solution6
29
Err-214
The READ? or MEAS? command is received when any trig-
ger source except for IMM is set.
[Trigger deadlock]
Solution6
30
Err-222
Entered value is outside the setting range or lacking for nec-
essary parameters
[Data out of range]
Solution6
31
Err-313
Calibration data is lost.
[Calibration memory lost]
Solution7
32
Err-314
Parameters saved by “*SAV” command are lost.
[Save/recall memory lost]
Solution8
33
Err-315
Saved parameters are lost.
[Configuration memory lost]
Solution6
34
Err-330
Self-test error
[Self-test failed]
Repair
35
Err-350
Error queue overflows.
[Queue overflow]
Solution9
No.
Error display
Error description
[ ]: Response to the ERR? command
Solution