7351 Series Digital Multimeter Operation Manual
7.1.4 DC Current Measurement (DCI)
7-5
7.1.4
DC Current Measurement (DCI)
Input
terminal
Range
Maximum display
Resolution
Resistance
between input
terminals
Measurement accuracy *1
(% of r digits)
Temperature
coefficient
(ppm of r
digits)/
C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST
MED
SLOW1, 2
AZ ON AZ OFF
mA-
COM
200 mA
199.99
199.999
10
A
1
A
0.5
or less
0.03+2
0.03+10
0.03+10
40+1
40+2
2000 mA
1999.9
1999.99 100
A
10
A
0.5
or less
0.05+2
0.05+7
0.05+6
50+0.7
50+1.5
10 A-
COM
10 A
10.999
10.9999
1 mA
100
A
0.06
or less
0.15+2
0.15+7
0.15+6
50+0.7
50+1.5
*1
For AZ ON. For AZ OFF, two digits are added to the digit section.
Maximum allowable
applied current
Input protection
Fuse replacement
Between mA and
COM terminals
2000 mA 2 A/250 V fast-blow fuse which is compliant
with IEC60127 sheet1
On the rear panel
Between 10 A and
COM terminals
10 A 15 A/250 V fast-blow fuse with 10000 A break-
ing capacity
Contact us to repair