7351 Series Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-35
3. Measurement condition command
Item
SCPI command
Description
Initial
value
Calculation
NULL calcula-
tion
[:SENSe]:<function>:REFerence
:STATe {ON,1|OFF,0} *1
ON, 1: ON
OFF, 0: OFF
[:SENSe]:<function>:REFerence
:STATe? *1
Reply: 1 or 0
[:SENSe]:<function>:REFerence
{<real>|MAXimum|MINimum} *1
NULL constant setting
<real>: -999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
Note:
The setting is disabled when the NULL calculation is
set to OFF.
(0)
[:SENSe]:<function>:REFerence?
{MAXimum|MINimum}
Reply:
d.dddddE
dd
Smoothing cal-
culation
:CALCulate:SMOothing {ON,1|OFF,0}
ON, 1: ON
OFF, 0: OFF
:CALCulate:SMOothing?
Reply: 1 or 0
:CALCulate:SMOothing:POINts
{<int>|MAXimum|MINimum}
Smoothing count
<int>: 2 to 100 (counts)
(10)
:CALCulate:SMOothing:POINts?
{MAXimum|MINimum}
Reply: ddd
Scaling calcula-
tion
:CALCulate:SCALing {ON,1|OFF,0}
ON, 1: ON
OFF, 0: OFF
:CALCulate:SCALing?
Reply: 1 or 0
:CALCulate:SCALing:A
{<real>|MAXimum|MINimum|
MEASurement}
Constant A (Zero cannot be set)
<real>:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(1)
:CALCulate:SCALing:B
{<real>|MAXimum|MINimum|
MEASurement}
Constant B
<real>:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(0)
:CALCulate:SCALing:C
{<real>|MAXimum|MINimum|
MEASurement}
Constant C
<real>:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(1)
:CALCulate:SCALing:A?
{MAXimum|MINimum}
Reply:
d.dddddE
dd
:CALCulate:SCALing:B?
{MAXimum|MINimum}
:CALCulate:SCALing:C?
{MAXimum|MINimum}
*1: <Function>: Specifies a function, which is not enclosed by double quotes, for "measurement function".