7351 Series Digital Multimeter Operation Manual
5.9.5 Comparator Calculation
5-26
5.9.5
Comparator Calculation
1. Function
The comparator calculation equation is shown below.
(Processing the special data)
• The overload (OL) of plus (+) data is judged as HI.
• The overload (OL) of minus (-) data is judged as LO.
The result whether HI, GO, or LO is displayed by setting COMP in
to ON.
The judgment result can be output to the display, buzzer, and status register.
(The buzzer sounds only when BEEP is set to ON.)
The comparator calculation can be applied to all measurement functions.
However, when MAX/MIN calculation is set to ON, the comparator calculation is applied to the value
before the MAX/MIN calculation.
2. Setting the judgment conditions
The setting examples of the High constant, Low constant, and buzzer are shown below.
HI = (Measurement value > HI setting value)
LO = (Measurement value < LOW setting value)
GO = (LOW setting value
Measurement value
HI setting value)
MENU
MENU
..3.. CALC
OFF/ON
COMP
MENU
..4.. CONST
High
Low
Changing the High constant setting
Changing the Low constant setting