10.0 CALIBRATION
ACTION DISPLAYS
Press the
[Tare/Zero]
key during
the self-test at power on. The scale
will ask to enter the password.
“ Pi n”
Default Password is 0000.
Enter “0” four times.
Password can be changed in
technical parameters.
Press
[Print]
.
“ Pi n” “ ----”
Select the scale to be set up by
using the
[Local/Remote]
key.
Press the
[Print]
key to enter the
technical section.
“tEch” “ LocAL” “ “
“tEch” “rEmo tE” “ “
Use the
[U.Wt.]
to select the
weighing unit to be used for setting
the calibration for the scale. The
arrow in the “
Weight
” window will
indicate the unit selected.
Press the
[Print]
key to continue.
“tEch” “ Uni t” “ “
You will enter the first parameter -
Calibration. Display will ask to
unload any weight on the platform.
Press the
[Print]
key to continue.
“Un L o Ad”
Enter the calibration weight to be
loaded and press
[Print]
.
Load the calibration weight onto
the scale and press
[Print]
again.
The Calibration is complete, the
scale will run the self-test during
which the weight should be
removed.
“SEL” “ 0010”
“Lo Ad”
© Adam Equipment Company 2005
28