Leveling the Balance
Aczet balances have a level control and adjustable leveling feet to compensate
for slight irregularities in the weighing bench surface. The balance is exactly
horizontal when the air bubble is in middle.
Turn the two leveling feet as desire picture in diagram so that air bubble comes in
middle.
Air bubble at
“12 o’clock”
Turn both leveling feet counter- clockwise.
Air bubble at
“3 o’clock”
Turn left leveling foot clockwise,
right leveling foot counterclockwise
Air bubble at
“6 o’clock”
Turn both leveling feet clockwise
Air bubble at
“9 o’clock”
Turn left leveling foot clockwise,
right leveling foot counterclockwise
Note
: The balance must be re-leveled each time it is moved to a new location.
Leveling Balances with a Weighing Capacity up to 10 kg
Leveling Balances with a Weighing Capacity over 10 kg
• Adjust the leveling feet until the air bubble is centered within the circle of the
level indicator
Connecting Electronic Peripheral Devices
Make absolutely sure to unplug the balance from DC power before you connect or
disconnect a peripheral device (printer or PC)to or from the interface port.
Warmup Time
To deliver exact results, balance must warm up to operating temperature for as leasted
below before the first weighing operation is carried out.
• CY (0.1mg) / CY (0.001ct) all Analytical models: at least 60 minutes
• CY (1mg) / CY (10mg) / CY (100mg) all Precision models: at least 30 minutes
Using Verified Balances as Legal Measuring Instruments in the EU* balance must warm
up for at least 24 hours after initial connection to DC power.