E.g. If the settings are as follows,
Factor = 1.02356
Accuracy = 0.00
LSD = 50

Now if 50 gm of weight is loaded on the pan the calculation for displayed weight will be as follows,
Weight * Factor
= 50 * 1.023456 
= 51.1728 

The displayed weight will be 51.150

The second digit after decimal point will change in multiples of 5 because,

Accuracy * LSD 
= 50 * 0.001
= 0.05

You can use the <TOGGLE> key at any time to switch the display between custom unit display, weighing unit 1 and weighing unit 
2. User can go to Standby mode by pressing on/off key and go to Simple Weighing with unit 1 by Cancel key and balance shows 
Current weights.

60

Printouts generated when Unit Toggling is done 
between Application Unit (

Cst

), Unit1 and 

Unit2.

Printouts generated when Unit Toggling  is done
 between Application Unit (

Cst

), Unit1 and 

Unit2.

When GLP ON

When GLP OFF

-----------------------------------
28-Jul-10        03:19PM
        Aczet
Model

CY 224

Ser.no.

9223102

Ver.no.         

r0.1.5.3

ID               

1234567

-----------------------------------

LID:

1111111

Factor      

3.023456    

Accuracy         

0.1    

Lsd                

2    

    +          

151.2 cst

    +

50.0000 g

    +

250.000 ct

Factor      

1.023456    

Accuracy       

0.001    

Lsd               

50    

    +         

61.400 cst

    +         

51.200 cst

-----------------------------------
28-Jul-10        

03:23PM

Name: 
        ................
-----------------------------------

Factor      

3.023456    

Accuracy         

0.1    

Lsd                

2    

    +          

151.2 cst

    +

50.0000 g

    +

250.000 ct

Factor      

1.023456    

Accuracy       

0.001    

Lsd               

50    

    +         

61.400 cst

    +         

51.200 cst

If  the balance is connected eternally  to PC or Printer through RS 232 C then, whenever user enter into the Custom Unit Menu & make 
changes  in the Factor, Accuracy and LSD setting, automatically printout is generated on the Peripheral attached.

In the printout, the new Factor, Accuracy and LSD values are printed. After this user can Press Print Key to Print the  reading on the display. 
Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.

Note : 

1) User cannot enter into the external calibration or menu when  GLP is ON 

           2) To print footer user will have to Press <CANCEL> key.

& footer has not been printed.

59

Содержание CY 1003

Страница 1: ...O p e r a t i n g M a n u a l Analytical Balances Carat Balances Precision Balances High Precision Industrial Balances This document hosted by www oldwillknottscales com...

Страница 2: ...Functions 44 6 1 Piece Counting 44 6 2 Percent Weighing 49 6 3 Custom Units 57 6 4 Animal Weighing 61 6 5 Formulation 65 6 6 Check weighing 69 6 7 Totalization 73 6 8 Density Determination 77 6 9 Pip...

Страница 3: ...safety standards you must comply with the provisions as specified in the applicable regulations for installation in your country Make sure that the voltage rating printed on the AC adapter is identic...

Страница 4: ...t Pan Cover CY 0 1mg 0 001ct 1 Pair of Corner hole Cover CY 0 1mg 1mg 0 001ct Cautionary notes Aczet balances may not be operated in hazardous areas Before attachment of the DC adapter check whether t...

Страница 5: ...feet 8 DC adapter socket 9 Provision for anti theft device 10 Spirit Level 11 RS232C interface 12 Additional Display Sockets 13 Foot Tare Switch Socket 14 Calibration Switch for Verified Balance Keys...

Страница 6: ...n on the bushes now place the Wind Shield over the pan such that it does not touch the weighing pan Note Ensure there must be gap between the pan and the Wind Shield Place the components listed below...

Страница 7: ...ce Aczet Balance are equipped with a lug for optional anti theft device The anti theft device cable with lock is suitable for all models It is available from ACZET under order number CAD01 2 1 Locatio...

Страница 8: ...time it is moved to a new location Leveling Balances with a Weighing Capacity up to 10 kg Leveling Balances with a Weighing Capacity over 10 kg Adjust the leveling feet until the air bubble is centere...

Страница 9: ...Print Key Press this key to send displayed values over the built in data interface to a DataPrint printer or a PC 7 Weight Units 8 Weight readout in the selected weight unit 9 Capacity Bar This indica...

Страница 10: ...form operating Instructions Press key briefly Press and hold key until desired display appears Automatic procedure Connect DC Adaptor and Power ON the balance It will display version number for softw...

Страница 11: ...mple Weighing 19 20 Simple Weighing 4 0 2 Tarring Place empty container on the balance The weight is displayed Press Tare key briefly the balance displays zero Add weighing sample to container the net...

Страница 12: ...following installation if required by national legislation 21 Switching of Weight Units 20 Jul 10 10 35AM Aczet Model CY 224 Ser no 9223102 Ver no r0 1 5 3 ID 1234567 LID 1111111 49 9999 g 249 9990 c...

Страница 13: ...d to the acceleration due to gravity at its location Calibration is necessary Before the balance is used for the first time At regular intervals during weighing operation After a change in location Pr...

Страница 14: ...done is message is displayed The adjustment can be terminated at any time using the CANCEL key The following message appears Abort 14 If Balance is connected externally to PC or Printer via Rs232 Cabl...

Страница 15: ...Et Release CAL key now The required calibration weight is shown on the display Place calibration weight in center of pan After the cal Test procedure is completed the difference between the actual th...

Страница 16: ...Calibration is perform and Cal done is displayed The adjustment can be terminated at any time using the CANCEL key The following message appears Abort If Balance is connected externally to PC or Prin...

Страница 17: ...until the desired setting appears Confirm settings Press TARE key breifly when desired setting appears Store settings Press and hold TARE Key until StoreED appears Abort Press CANCEL Key briefly The b...

Страница 18: ...y user will remain unchanged Above instructions are to be followed for all menu options except Reset 33 Application Menu 5 2 1 Special applications and functions Your balance can do more than just wei...

Страница 19: ...ve no function available in the weighing mode factory setting Note Above function will replace preset unit 3 automatically 35 36 5 2 2 Unit 1 2 3 selecting The following weight units can be selected W...

Страница 20: ...to the External calibration CAL OFF When user press hold CAL key directly Menu appears on the display without CAL Et option Thus user cannot enter into the calibration process CAL INT If the user sele...

Страница 21: ...enu block you tell the balance how a value should be transferred to a peripheral device e g computer Prn req The next possible stable value will be transferred after triggering of the Print key Prn Co...

Страница 22: ...ing GLP off If the user select this option then the balance print format are not compliance to ISO GLP GMP GLP on If the user select this option then the balance print format are compliance to ISO GLP...

Страница 23: ...t reference no as the reference number You can accept this suggestion or select one of the other reference numbers available 5 10 20 50 100 pieces Free wref Update by briefly pressing the TOGGLE key N...

Страница 24: ...will be with respect to the reference calculation based upon the FREE setting 45 6 1 2 wRef settings If the piece weight is known it can be entered directly To do this press the TARE key when the syst...

Страница 25: ...Y 224 Ser no 9930508 Ver no r0 1 04 ID 1234567 LID 1000000 NRef 5 pcs WRef 4 00 g Qnt 5 pcs NRef 50 pcs WRef 0 40 g Qnt 50 pcs Qnt 25 pcs NRef 5 pcs WRef 2 00 g Qnt 5 pcs 29 Jul 09 03 47PM Name Printo...

Страница 26: ...Percent Weighing You can use the TOGGLE key at any time to switch the display between the Percent display weighing unit 1 and weighing unit 2 Note The current set weight remains stored until it has be...

Страница 27: ...nit through Toggle key Press Print Key to get the print out of those corresponding units Printouts generated when Unit Toggling is done between Application Unit Unit1 and Unit2 and Reference Weight is...

Страница 28: ...WW 100R Residue The dry content of the sample is displayed and printed out as a percentage of the wet weight ww initial weight 100 When the results are printed out the dry content is designated 100R R...

Страница 29: ...nt measured value on display mode is greater or less than the predefined limit value i e greater than 999 99 or less than 999 99 the balance displays Over range Or 55 100L Loss Content Treat the sampl...

Страница 30: ...hift the decimal position in a cyclic way After proper setting of factor press the TARE key Note Please refer error conditions for errors occurred in storing the Factor Accuracy and LSD settings Accur...

Страница 31: ...ul 10 03 19PM Aczet Model CY 224 Ser no 9223102 Ver no r0 1 5 3 ID 1234567 LID 1111111 Factor 3 023456 Accuracy 0 1 Lsd 2 151 2 cst 50 0000 g 250 000 ct Factor 1 023456 Accuracy 0 001 Lsd 50 61 400 cs...

Страница 32: ...eighing Manual Animal Weighing presupposes that you have selected the F Anl in the user menu Procedure Place the empty container on the pan Press the TARE key briefly to tare the balance Press the TOG...

Страница 33: ...time to switch the display between animal weighing weighing unit 1 and weighing unit 2 User can go to Standby mode by pressing on off key and go to Simple Weighing with unit 1 by Cancel key and balan...

Страница 34: ...owing output addition of 20 0070 gm N2 20 0070 g Tot 39 0038 g To view the total weight press the CAL and TOGGLE key together The print command is given automatically N 2 Tot 39 0038 g 66 6 5 Formulat...

Страница 35: ...ough Toggle key Press Print Key to get the print out of those corresponding units 6 5 2 Auto Formulation Auto formulation presupposes that you have selected the F Form AUTO in the user menu Place the...

Страница 36: ...target and tolerance values If entering the check weighing for the first time the target and tolerance setting menu will be displayed Target and Tolerance Settings Press and hold the TOGGLE until the...

Страница 37: ...se corresponding units Note 1 User cannot enter into the external calibration or menu when GLP is ON 2 To print footer user will have to Press CANCEL key footer has not been printed Load the weight on...

Страница 38: ...d by pressing CANCEL key Note Your balance suggests the last set reference no as the reference number You can accept this suggestion or select one of the other reference numbers available 5 10 20 50 1...

Страница 39: ...nce now needs the weight of a reference number Press and hold the TOGGLE key until you are prompted to load the reference pieces Your balance suggests the last selected reference number You can accept...

Страница 40: ...termined on the basis of density The weighing unit of density determination will be grams Density determination presupposes that you have selected the F Den in the user menu To calculate the density o...

Страница 41: ...weight of sample in liquid With Wt Liqd flashing on the display every 10 seconds Now immerse the sample in the liquid When the weight is stable press the CAL key to accept the weight of sample in liqu...

Страница 42: ...23 1 ct 18 9350 g 94 675 ct Mode GOLD Liquid WATER Temperature 25 0 Pur 23 1 ct 18 9350 g 94 675 ct If the balance is connected externally to PC or Printer through RS 232 C then whenever user enter in...

Страница 43: ...default temperature is 25 0 C Pressure Settings Press TOGGLE key to enter F PiP settings Select PrES in F PiP settings by pressing TOGGLE key Press TARE to enter the PrES settings The user can select...

Страница 44: ...iP settings by pressing TOGGLE key 0 Press TARE to enter the V settings 0 The user can change the volume with the help of CAL MENU key and TOGGLE key Pressing the CAL MENU key will shift the cursor to...

Страница 45: ...Es 10005 95 uL Es 100059 47 Sr 0 34 uL CV 0 00 Results V1 2 1 15023 82 uL 2 15023 62 uL 3 15023 62 uL 4 15023 62 uL 5 15023 72 uL V1 2 11 00 uL Va 15023 68 uL Es 15012 68 uL Es 136478 92 Sr 0 09 uL C...

Страница 46: ...ce the empty container on the pan Press the TARE key briefly to tare the balance Press the TOGGLE key for entering the F StAt function Note To use the F StAt function it is assumed that the user has a...

Страница 47: ...it gets deleted Note The minimum weight on the pan should be 10d where d is the accuracy of the scale if less than that then the weight will not be taken into consideration for calculating statistics...

Страница 48: ...t menu Auto Cal Menu Confirm Setting Press briefly Press Hold Increments the value of digit Change Main menu options Shifts the digit from left to right Quit the Current Parameter Menu ID LID SETTINGS...

Страница 49: ...ings They are TIME In this submenu user can set the time in hours minutes seconds AM PM 24 hrs DATE In this submenu user can set the date Month Years Operating Instructions Press the PRINT key for 2 s...

Страница 50: ...ating Instructions Press the PRINT key briefly when coming out from stand by or Power on mode 7 2 1 SET ID Last stored ID is displayed with first digit flashing Flashing digit indicates that digit val...

Страница 51: ...ternal Calibration can be triggered automatically by the following Time Temperature Power on Calibration Operating Instruction Press the CAL key briefly when coming out from stand by or Power on mode...

Страница 52: ...ll Enter control panel Open ACCESSIBILITY OPTIONS from control panel In the general tab turn on serial key option Set the baud rate and COM port from the settings option 7 3 3 Power on Calibration Pre...

Страница 53: ...indows User can set Unit option as OFF Only the numerical value will be sent to windows and not the unit Windows Separator settings Select Windows print option as ON and press the TARE key now press t...

Страница 54: ...re Version ID 1234567 ID Dotted line LID 1111111 Lot ID nRef 170 pcs Counting Reference Sample Quantity wRef 0 2945 g Counting Reference Weight Qnt 170 pcs Counting Result 50 0650 g Weighing Result 25...

Страница 55: ...e Feed HH Unit Symbol 107 Purpose Your balance is equipped with an interface port for connection to a computer or other peripheral device You can use an on line computer to change start and or monitor...

Страница 56: ...a information between Bi directional communications it shows end of command frame Commands W If host computer send this command to balance then balance will Send weight with current unit T If host com...

Страница 57: ...n the tolerance limit GLP is ON and user tries to enter in to the User Menu before the footer is printed Calibration Display shows any weight other than 0 00 and user tries to Calibrate the balance In...

Страница 58: ...e Regular servicing by a ACZET technician will extend the service life of your balance and ensure its continued weighing accuracy ACZET can offer you service contracts with your choice of regular main...

Страница 59: ...dditional protection Do not use stainless steel cleaning agents that contain soda lye caustic acetic acid hydrochloric acid sulfuric acid or citric acid The use of scrubbing sponges made of steel wool...

Страница 60: ...modification by other than ACZET In lieu of a properly returned warranty registration card the warranty period shall begin on the date of shipment to the authorized dealer No other express or implied...

Страница 61: ...mg Series Balances Draft Shield High Carat Balances CY 0 001ct Series Balances Draft Shield Low Precision Balances CY 1mg Series Balances Draft Shield Low Top Loading Precision Balances CY 10mg Series...

Страница 62: ...apacity Precision Balances CY 100mg Series Balances High Capacity Precision Balances CY 1mg Series Balance with wind shield 402 419 18 129 95 247 347 268 05 190 5 212 171 5 204 5 189 5 159 81 67 5 231...

Страница 63: ...mg g mg mg s 0 C 0 C g kg mm mm mm V Hz VA Analytical Balances 13 Specifications 123 CY 54C 51 0 1 1 51 0 1 0 2 II 15 to 30 C Internal 25 E2 124 Model Weighing Capacity Readability d Accuracy e Tare R...

Страница 64: ...Hz VA Precision Balances CY 1003 1000 0 001 0 01 1000 0 002 0 003 I 18 to 30 C External 500 F1 CY 363C 360 0 001 0 01 360 0 001 0 002 II 15 to 30 C Internal 200 F1 126 Model Weighing Capacity Readabi...

Страница 65: ...Weighing Class Response time average Operating temperature range Calibration External cal wt of at least accuracy class Net Weight approx Pan size Dimensions W x D x H DC power source Power requiremen...

Страница 66: ...e gram kilogram pound ounce troy ounce grain pennyweight carat Milligram momme mesghal Hong Kong tales Singapore taels Taiwan tales baht g g g g g g s 0 C g kg mm mm V Hz VA Top Loading Precision Bala...

Страница 67: ...e taels Taiwan tales baht Kg g g Kg g g s 0 C Kg kg mm mm V Hz VA High Capacity Precision Balances 132 CY 50001H 50 1 10 50 1 2 III 15 to 45 C External 25 F1 Model Weighing Capacity Readability d Accu...

Страница 68: ...ct ct ct ct ct ct s 0 C kg mm mm mm V Hz Carat Balances 134 Model Weighing Capacity Readability d Accuracy e Tare Range Subtractive Repeatability std deviation Linearity Weighing Class Response time...

Страница 69: ...supplied with the instruments 14 Accessories Option Statistical Printer CPR 02 with Date Time Statistics Remote Display SRD01 Calibration Weights F1 ERTL F2 with certificate for further details conta...

Отзывы: