If Balance is connected externally to PC or Data Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated
after the completion of Calibration process.
Calibration test determines the difference between the actual weight and the measured weight
Calibration test can be turned ON or OFF from the user menu.
When ON, cal test would be performed on external or internal calibration whichever is selected in
User Menu.
Procedure
ð
Have required calibration weight ready
ð
Press and hold <CAL> key, display, shows “CAL Et”
ð
Release <CAL> key now.
ð
The required calibration weight is shown on the display.
ð
Place calibration weight in center of pan.
ð
After the cal Test procedure is completed the difference between the actual & the
measured weight will be displayed on display.
The adjustment can be terminated at any time using the <CANCEL> key.
The following message appears : ‘Abort’
Calibration Test Report
28
27
Calibration (adjusting)
When GLP ON
--------------------------------
14-Jul-10
00:03AM
Aczet
Model
CY 224
Ser.no.
Ver.no.
ID
--------------------------------
Calibration:
External
W-ID ..................
Temperature
Set
Diff.
Calibration Test Done
---------------------------------
14-Jul-10
00:03AM
Name: ................
---------------------------------
1111111
r0.1.04.01
860054081
30.710'C
+200.0000g
+ 0.0047g
When GLP OFF
Calibration:
External
W-ID ..................
Temperature
Set
Diff.
Calibration Test Done
30.710'C
+200.0000g
+ 0.0047g
4.3 Calibration Test