6.1.3 Updating Settings
The Update feature improves the precision of piece counting results.
The average piece weight (reference) is recalculated with each reference
optimization. Because the new pieces that have been placed in the weighing pan
increase the basis for the calculation, the references, and therefore the result of the
piece count, are more exact.
Select the UPDATE feature from the reference menu. The reference can be updated
by pressing the CAL key which is confirmed by the displaying 'UPDATED' on the display.
Note
• The number of pieces placed in the weighing pan is larger than the reference piece
number shown on the display.
• The number of pieces placed in the weighing pan is not greater than twice the most
recently saved reference piece number (e.g. If the display shows 100 pcs the
added pieces should not be greater than 200).
• Update feature cannot be selected if the previous selected reference was 'wRef'
47
Printouts generated when Unit Toggling is done
between Application Unit (
Pcs
), Unit1 and
Unit2 and Reference Weight is changed
When GLP ON
------------------------------------
29-Jul-09
03:46PM
Aczet
Model
CY 224
Ser.no.
9930508
Ver.no.
r0.1.04
ID
1234567
------------------------------------
LID:
1000000
NRef
5 pcs
WRef
4.00 g
Qnt +
5 pcs
NRef
50 pcs
WRef
0.40 g
Qnt +
50 pcs
Qnt +
25 pcs
NRef
5 pcs
WRef
2.00 g
Qnt +
5 pcs
------------------------------------
29-Jul-09
03:47PM
Name: ................
------------------------------------
Printouts generated when Unit Toggling is done
between Application Unit (
Pcs
), Unit1 and
Unit2 and Reference Weight is changed
When GLP OFF
nRef
5 pcs
wRef
2.00 g
Qnt +
5 pcs
Qnt +
10 pcs
nRef
10 pcs
wRef
2.00 g
Qnt +
10 pcs
Qnt +
5 pcs
nRef
50 pcs
wRef
0.20 g
Qnt +
50 pcs
Qnt +
100 pcs
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the reference menu of Piece Counting
function & make changes in the reference setting, automatically printout is generated on the Peripheral attached.
In the printout, reference number “nRef” and reference weight “wRef” is printed. After this user can Press Print Key to Print the reading on the display.
Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
48
Note :
1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
Printout: Counting
nRef
+
10
:
Reference sample quantity
wRef
+
21.14 g :
Reference weight
i.e. weight of one piece
Qnt
+
500 pcs :
Calculated quantity