Percentage Weighing (%) (in 100R / 100L / AtroM / AtroD)
Percent Weighing (%) presupposes that you have preselected the “F per” function in
the menu
Procedure
ð
Press the <TARE> key to tare the balance.
ð
Press the <TOGGLE> key briefly until “PER (%)” appears on the display.
Your balance now needs the weight of a reference percent (%).
ð
Press and hold the <TOGGLE> key until you are prompted to load the reference
PER (%).
ð
Your balance suggests the last set reference % as the reference percent (%)
Press the <TOGGLE> key until the following option is displayed (100R / 100L / AtroM
/ AtroD)
ð
Now place reference sample on the pan which is to be analyzed (Initial Weight).
ð
Then press <TARE> key. Until horizontal dashes are displayed, your balance is
calculating the reference.
ð
After your balance has determined the reference weight, it is ready for Percent
Weighing in Differential weighing.
ð
Now treat the sample which includes process like drying, centrifugation, filtering,
incineration, vaporization, coating, etc. After the sample has been processed, re-
weigh it (residual weight). The balance then determines the difference between
the two weighed values.
You can use the <TOGGLE> key at any time to switch the display between the Percent
(%) display, weighing unit 1 and weighing unit 2.
Note
: If the current measured value on display mode is greater or less than the
predefined limit value (i.e. greater than 999.99 % or less than –999.99 %) the balance
displays Over range (
----Or----
)
55
100L% (Loss Content)
(Treat the sample& Re-Weigh it)
Printouts generated when Unit Toggling is done
between Application Unit (
Per %
), Unit1 and
Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done
between Application Unit (
Per %
), Unit1 and
Unit2 and Reference Weight is changed
When GLP ON
When GLP OFF
-----------------------------------
28-Jul-10 03:19PM
Aczet
Model
CY 224
Ser.no.
9223102
Ver.no.
r0.1.5.3
ID
1234567
-----------------------------------
LID:
1111111
pRef
ATROD %
wRef
0.5000 g
Pct +
100.00 %
Pct +
500.00 %
+
50.0000 g
+
250.0000 ct
-----------------------------------
28-Jul-10
03:23PM
Name:
................
-----------------------------------
pRef
ATROD %
wRef
0.5000 g
Pct +
100.00 %
Pct +
500.00 %
+
50.0000 g
+
250.0000 ct
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the reference menu of Percent
Weighing function & make changes in the reference setting, automatically printout is generated on the Peripheral attached.
In the printout, reference percent “pRef” and reference weight “wRef” is printed. After this user can Press Print Key to Print the reading on
the display.
Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
Note :
1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.