To correct the weighing results, the TARE key need to be pressed when the difference
is display upon pressing the TARE key. Actual calibration is performed ‘CAL done’ is
displayed and the weighing results are corrected as shown alongside.
Procedure
ð
Have required calibration weight ready
ð
Press and hold <CAL> key, display, shows “CAL Et”
ð
Release <CAL> key now.
ð
The required calibration weight is shown on the display.
ð
Place calibration weight in center of pan.
ð
The difference between the actual & the measured weight will be displayed.
ð
Press the Tare key when the difference is displayed.
ð
Actual Calibration is perform and Cal done is displayed.
The adjustment can be terminated at any time using the <CANCEL> key.
The following message appears : ‘Abort’
If Balance is connected externally to PC or Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated after
the completion of Calibration process.
Calibration Test Report
30
29
Calibration (adjusting)
Calibration
:
External
W-ID
............
Temperature
32.905'C
Set
+ 200.0000g
Diff.
-0.1235g
External Cal Done
Diff.
0.0000g
------------------------------------
14-Jul-10
03:46PM
Aczet
Model
CY 224
Ser.no.
9930508
Ver.no.
R0.1.04
ID
1234567
------------------------------------
Calibration:
External
W-ID ..................
Temperature
32.898'C
Set
+ 200.0000g
Diff.
-0.1235g
External Cal Done
Diff.
0.0000g
------------------------------------
14-Jul-09
03:46PM
Name: ................
------------------------------------
When GLP OFF
When GLP ON
4.4 Calibration Test with Actual Calibration