Appendix 1 Meter Specifications
Feature
Specifications
Methodology
Reflectance Photometer
Test Time
≤5 min
Measurement Range
UA: 0.090mmol/L-1.200mmol/L(1.51mg/dL-20.17mg/dL)
CR: 0.044mmol/L-1.320mmol/L(0.50mg/dL-14.93mg/dL)
UR: 0.90mmol/L-40.00mmol/L(5.41mg/dL-240.24mg/dL)
Specimen
Whole blood(capillary blood, venous whole blood), plasma
and serum
Specimen Volume
35 μL or 10 μL
Power Source
Charging battery;
Note: during charging, the meter is non-operational
Units of Measurement
mg/dL or mmol/L
Memory
500 records
Automatic Shut Off
After 5 minutes of inactivity
Meter Size
135 mm × 66mm × 19 mm
Weight
90g
Meter storage and transportation t
conditions
0°C - 55°C; ≤ 90% RH
System Operating Conditions
10°C - 35°C; ≤ 80% RH; indoor only
pollution degree of the intended
environment
2
System operating altitude
Max. 2000 m
Test strip storage and transportation
t conditions
2°C - 30°C; ≤ 90% RH
Optical Verify Strip storage and
transportation t conditions
2°C - 30°C; ≤ 90% RH
Meter Software Version Number
1.0