03. Compare the code number on the code chip with the code number printed on
the test strip pouch label and ensure the two numbers are identical to avoid
inaccurate results.
04. Press and hold
for 2 seconds to enter the Setting mode. And press ► to
change the selection. Please press to enter the
CTR
. Press ◄ or ► to
switch between
OPt
and
SOLU
. Press to confirm it and enter the
SOLU
after selected.
05.
The meter will flash the test strip icon. Insert a test strip completely into the
Test Strip Holder in the same direction as the arrows printed on the strip until it
cannot be inserted any further.
06.
When the meter is flashing the blood drop symbol, open the screw cap of the
control solution bottle and turn the bottle upside down. Squeeze the control
solution bottle gently and discard the first drop. If there are bubbles in the
previous drop, squeeze the bottle and discard another drop until there are no
bubbles in the drop. Apply the next drop to the specimen well on the test strip
while keeping the bottle vertically upside down. Use about 35μL of control
solution for the test strip. Make sure the control solution is applied directly into
the specimen well and that there is no bubble in the solution drop.
Note:
• Make sure the bottle is completely vertical when applying the solution to the
strip. The volume will be inconsistent if the bottle is not completely vertical.
• Gently squeeze the solution completely onto the tip of the bottle and drop it
freely into the sample well. Avoid touching the strip with the tip of the bottle.
07. For the test, two kinds of control solutions need to be tested on two separate
test strips. Remember to switch to a new test strip after the control solution has
been tested on the first strip.