Section 9 Quality Control
Each lab should use its own standards and procedures for performance. Test known
specimens/controls at each of the following events in accordance with local, state,
and/or federal regulations or accreditation requirements:
• When a new package of test strips is opened
• When a new operator uses the meter
• When test results seem inaccurate
• After performing maintenance or service on the meter
If quality control tests do not provide expected results, perform the following checks:
• Ensure that the test strips used are not expired.
• Ensure that the test strips are fresh from a new canister or package.
• Ensure that the controls are not expired.
• Repeat the test to ensure no errors were made during the test.
Control Solution Testing
Renal Function Control Solution testing is performed in a very similar manner to
blood tests. The Renal Function Control Solution is used instead of blood.
Note:
Make sure the control solution and all the test materials reach operating
temperatures of 10-35°C prior to testing. Tests can only be accurately performed
when the control solutions and test materials are within this temperature range.
01. Turn on the meter.
02. Insert the code chip into the meter. Refer to Coding the Meter in the
instructions for use for details. Make sure the test strip box is firmly capped
and the control solution is tightly closed before use.