discolored, or damaged optical verify strip.
• Do not use it after the expiration date.
• For in vitro diagnostic use only.
Storage and Handling
• Store optical verify strips in a cool, dry place. Store away from heat and direct
sunlight.
• Transport and store in its closed canister within 2-30°C with less than 90%
humidity.
• Replace the cap on the strip canister immediately after removing a strip.
Expired strip may produce incorrect test results.
Control Solution
The Renal Function Control Solution contains stabilizers, preservatives and added
chemicals. To confirm that the test strip and meter are working together properly
and that the test is being performed correctly, the control solution is applied to the
specimen well of a test strip that has been inserted in the meter. Refer to the "CTR"
section in the instructions for use for more information.
Note:
The Renal Function Control Solution is intended for validating Renal
Function testing while using the Renal Function Monitoring System. Both levels
of control solutions must be tested and fall within the assigned values printed on
the bottles.
Refer to the control solution package insert before using the controls. The control
solution bottle is labeled with the acceptable range that is specific for that lot of
control solution. The system is working properly if the control value displayed by the
meter is within the acceptable range printed on the bottle label. If the value does
not fall within the range, refer to the Control Solution Package Insert for further
instructions.