Section 4 Initial Setup
Before testing, ensure the following procedures are followed.
Turn on Meter
Press to turn the meter on.
The meter will automatically turn off after 5 minutes of inactivity.
Coding the Meter
Each time a new box of test strips is used, the new code chip included in the box
must be inserted into the meter. Compare the code number on the code chip with
the code number printed on the test strip box or the foil pouch. Results may be
inaccurate if the two numbers are not identical. Insert the new code chip into the
code chip slot of the meter. It should easily snap into place. The code chip should
remain in the meter. Do not take it out until a new box of test strips is needed. The
code number will appear on the Initial Screen after startup.
If the code chip is not properly inserted into the code chip slot or if it is missing, the
meter will display E-6.