24
Versions 1.0
Note:
For testing capillary blood, use the second drop of blood for accurate
results.
The meter will begin testing automatically with
four dashes
in a line flashing on the
display indicating the test is in progress.
Results will be displayed within 300 seconds.
Note:
The date in the display will be shown according to the mode you
previously selected.
Remove the used test strip. The meter will return to the Initial Screen and is ready
for another test strip to be inserted and to perform a new test.
Note: Discard all blood specimens, used Test strips, and materials
carefully� Treat all blood specimens as if they were infectious material�
Follow proper precautions and obey all local regulations when discarding
blood specimens and materials�
Perform daily cleaning when testing is completed for the day. Refer to the
Maintenance section.
The meter will automatically turn off after 5 minutes of inactivity or when is
pressed.
RFM-101 insert 20191125.indd 24
2019/11/27 10:42:07