GX-A/GF-A Maintenance manual
Page 47
6.12 Half-initialization
Half-initialization changes all the user settings(*1) to the default values. Follow the
procedure below.
Step 1. Display the function setting mode
Unit
,
after getting into check mode shown
on page 29.
Step 2. Press the SAMPLE key several times.
initHf
is displayed.
Step 3. Press the PRINT key.
Hf
is displayed. (To cancel the operation, press the
CAL key or PRINT key.)
Step 4. Press the RE-ZERO key.
Hf
is displayed. (To cancel the operation, press
the CAL key.)
Step 5. Press the PRINT key.
<
Hf
and then
end
is displayed. The
Half-initialization is completed.
(*1) The user settings
・
Function setting
・
Value of the external mass for calibration
・
Content of data memory function
・
Density data of liquid for density measurement
・
Password
No
G
o
G
o