GX-A/GF-A Maintenance manual
Page 12
Problem
Cause
Check
Solution
Type
Unstable
weighing
data,
repeatability
error
Force motor,
Analog
board
assembly
Check the operation of weight /
temperature offset A/D.
(Refer
to
“6-13
Method
of
identifying defect location
”)
Replace
the
analog
board
assembly
E
Then if it is OK, the cause will be
the force motor.
Check the force motor
A
Pan assembly
Check that the pan assembly is
correctly installed
Install
the
pan
assembly
correctly.
F
Check for foreign matter between
the pan and dust plate.
Clean the area around the pan
assembly and the dust plate.
F
Check that the dust plate does not
touch the pan assembly.
Install the dust plate correctly so
that it does not touch the pan.
F
Damage
in
flexures
Check the condition of tension and
fulcrum flexures and lever.
Replace
the
parts
and
reassemble the force motor.
A
Magnet
assembly
Check for dust particles between
the magnet and the force coil.
Clean the force motor assembly.
(Refer
to
“5. Force Motor
Disassembly and Reassembly
”)
A
Force
motor
assembly
Check that the flexures are in good
conditions
and
are
correctly
installed.
Repair the force motor assembly.
(Refer
to
“5. Force Motor
Disassembly and Reassembly
”)
A
Corner load
error
Force
motor
assembly
Check that the flexures are in good
conditions
and
are
correctly
installed.
Perform corner load adjustment.
F
If it does not work well,
disassemble and reassemble the
balance.
(Refer
to
“5. Force Motor
Disassembly and Reassembly
”)
A
Hysteresis
error
Tension
or
Fulcrum flexures
Check the condition of tension and
fulcrum flexures for distortion.
Replace with substitute items.
Disassemble and reassemble the
balance.
(Refer
to
“5. Force Motor
Disassembly and Reassembly
”)
A
Linearity
error
Force
motor
assembly
Follow the linearity check
procedure.
Input the linearity data.
Refer to page 40.
F
If it does not work well,
disassemble and reassemble the
balance.
(Refer
to
“5. Force Motor
Disassembly and Reassembly
”)
A
After
calibrated
using
the
internal
mass,
the
weighing
data is not
correct
(OnlyGX-A)
Force
motor
assembly
After auto calibration using the
internal mass, check the value
using the specified mass.
Fine adjustment of the linearity.
Refer to page 40~45.
F