GX-A/GF-A Maintenance manual
Page 31
6.5 Initialization
<CAUTION>
· Do not proceed with this initialization if you cannot prepare a
temperature-controlled room. (It is necessary to adjust temperature data
after this initialization)
Note that the following data is initialized by this operation.
• Temperature compensation data for each unit (Then some fixed data is input
compulsorily)
• Linearity compensation data (The data is cleared), serial number, weighing
units
• Internal parameter settings (
Certain
values are input compulsorily)
Initialization procedure
Follow the procedure below to completely initialize the data stored in the non-volatile
memory.
Step 1. See the check mode menu flowchart on page 29 to display basic setting mode
type
or
apr
.
Step 2. Press the SAMPLE key several times.
init
will be displayed.
Step 3. Press the PRINT key
.
Aall
will be displayed. To cancel the operation, press the
CAL or PRINT key. The next item will be displayed.
Step 4. Press the RE-ZERO key.
Aall
will be displayed. To cancel the operation, press
the CAL key. The next item will be displayed
Step 5. Press the PRINT key. <
Aall
, then
end
will be displayed. The next item will be
displayed. Change the other items as necessary.
Step 6. To quit the operation, press the ON/OFF key. The display will be turned OFF.