GX-A/GF-A Maintenance manual
Page 32
6.6 Model type Selection (Only for maintenance electric board)
Model type selection is available for the maintenance board only.
Model selection procedure
Follow the procedure below to select the balance model type.
Step 1. Verify the following item of the 7PZ-6594B/C (Main board
for the maintenance
).
□
Verify the type of the board
・
For each type (B/C), verify that the jumper setting is correct. (Refer to the
Technical information
“3. Circuit Diagram” and “4. Part Layout”)
□
For the main board unit, verify if it is adjusted completely
・
Connect the AC adapter to the main board unit. If
error 9
is displayed, the
main board is not adjusted yet and it should be adjusted for a maintenance
board.
Step 2. See the check mode menu flowchart on page 29 to display basic setting mode
typeE
.
Step 3. Press the PRINT key. Confirm the type on the display as follows.
Type of board
Model type which can be set
Type displayed initially
7PZ-6594B
GX-203A/303A/403A/603A/1003A/1603A
GF-123A/203A/303A/403A/603A/1003A/1603A
GX-1603A
7PZ-6594C
GX-2002A/3002A/4002A/6002A/10002A
-6001A/10001A
GF-1202A/2002A/3002A/4002A/6002A/10002A
-6001A/10001A
GX-10002A
Use the following keys to select the model type
RE-ZERO key: changes the model type. (Change the model type according to
the kinds of PZ)
PRINT key:
saves the final model type.
CAL key:
cancels the data.
When the model after change is the same as that before change, the stabilization
indicator illuminates.
Step 4. Press the PRINT key.
end
appears. Then, the next item appears.