GX-A/GF-A Maintenance manual
Page 7
3.2 Test Details
External mass calibration (GF-A series)
After external mass calibration by pressing CAL key, place the specified mass on the pan and
read the displayed value. Verify that the difference between the maximum value and the
minimum value is within the specifications.
GF-A series
Model
Masses
Specifications
123
100g
±0.001g
203
200g
303
300g
403
400g
603
600g
1003
1000g
1603
1600g
±0.002g
1202
1kg
±0.02g
2002
2kg
±0.01 g
3002
3kg
4002
4kg
6002
6kg
10002
10kg
±0.02g
6001
6kg
±0.1 g
10001
10kg
Internal mass repeatability (GX-A series)
Calibrate using the internal mass
.
Perform this three times,
verify
that the all data be within the specifications.
GX-A series
Model
Masses
Specifications
203
200g
±0.010 g
303
300g
403
400g
603
600g
1003
1000g
1603
2002
2kg
±0.10g
3002
3kg
±0.15g
4002
4kg
6002
5kg
10002
6001
5kg
±0.5g
10001