GX-A/GF-A Maintenance manual
Page 26
6. Adjustments
6.1 Adjustment Flow Chart
Force motor assembly
Initial setting (initialization, model type selection...if necessary)
Coarse operation check (A/D operation)
Coarse adjustment (calibration and corner load adjustment)
Temperature adjustment (enter high and low temperature data)
Fine adjustment (corner load adjustment, linearity, hysteresis)
Internal parameter setting (selection of weighing units and serial number setting)
Performance test (corner load adjustment, repeatability, linearity)
Check the maintenance main board