10.06 Other Error Screens
There is no available memory within the instrument to store further results. Press ‘
OK
’, upload all stored results to
the PC and try again.
There is an internal fault with the chamber or connections to the chamber. Contact 3M or local 3M distributor.
Error Message
User Action
Unexpected error:
requesting results
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or
remove from the docking station, then reconnect and upload the results again.
Unexpected error:
authenticating
Confirm user ID and passwords are correct, disconnect the 3M™ Clean-Trace™
NGi luminometer from the USB lead, or remove from the docking station, then
reconnect and upload the results again.
Unexpected error:
updating password
#Reason
Confirm user ID and passwords are correct, disconnect the 3M™ Clean-Trace™
NGi luminometer from the USB lead, or remove from the docking station, then
reconnect and upload the results again.
Unexpected error:
uploading results
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or
remove from the docking station, then reconnect and upload the results again.
Not all sample plans
have been downloaded -
Error when deleting and
downloading sample plans
Using the 3M™ Clean-Trace™ Online Software administration website, check
that sample plans have been assigned to the 3M™ Clean-Trace™ NGi
luminometer user.
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or
remove from the docking station, then reconnect and upload the results again.
Not all sample plans have
been downloaded
#Please Retry Upload
Sequence
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or
remove from the docking station, then reconnect and upload the results again.
There are no Sample Plans
Available
Using the 3M™ Clean-Trace™ Online Software administration website, check
that sample plans have been assigned to the 3M™ Clean-Trace™ NGi
luminometer user.
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or
remove from the docking station, then reconnect and upload the results again.
Error received:
deleting results
#results have not been
deleted
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or
remove from the docking station, then reconnect and upload the results again.