8.0 USER GUIDELINES
WARNING:
Do not service the 3M™ Clean-Trace™ NGi Luminometer, Docking Station or power supply.
Use only 3M authorized service personnel.
WARNING:
Do not use the 3M™ Clean-Trace™ NGi Luminometer or power supply if the enclosure is
broken.
i The 3M™ Clean-Trace™ NGi Luminometer is designed to be robust and easy to use but it is a sensitive
instrument which should be handled with care.
ii The sample chamber should be kept closed or covered by the cap between readings.
iii Take care not to contaminate the inside of the chamber by spillage of liquid sample or inserting a swab without
returning it to the device.
iv Always remove to the sample before switching
OFF
.
v Always switch
OFF
after use.
vi
WARNING:
do not use a power supply other than the one specified by 3M.
vii Reagent test kits should be stored and used in accordance with the instructions provided with the kit.
viii Use only reagents supplied by 3M or local 3M distributor.
ix Results are independent of temperature over the range of 15-25°C. A lower reading will be seen at
temperatures below 15°C. However, if tests are always performed at this lower temperature, consistent results
can be obtained.
x The sample chamber is designed to prevent access of ambient light during measurement but the instrument
has an extremely sensitive detector so some care must be exercised. It is prudent to avoid use of the
instrument in bright, direct sunlight.
xi Do not leave the 3M™ Clean-Trace™ NGi Luminometer in a location exposed to steam, high humidity, dust or
vibration.
xii Remove the batteries when the 3M™ Clean-Trace™ NGi Luminometer will not be used for an extended
period.
Read the Test Kit Instructions for complete information specific to the type of test.