Bair Paws Model 875 Warming Unit Service Manual
26
Check the 3M
TM
Bair Paws
TM
system website to ensure you have the most recent version of this document.
www.bairpaws.com reorder #202036A
Safety System Characteristics
thermostat
Independent electronic and heater (electromechanical)
overcurrent protection
Dual input fused line�
safety feature
Over-temperature protection: color indicator light
illuminates, audible alarm sounds, and heater shuts down
(blower continues to run)�
alarm system
Fault: amber Fault indication flashes, alarm sounds�
certifications
IEC 60601-1; IEC/EN 60601-1-2; UL 60601-1; CAN/CSA-
C22�2, No�601�1; EN 55011; EN 80601-2-35�
classifications
MEDICAL — GENERAL MEDICAL
EQUIPMENT AS TO ELECTRICAL SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2,
No.601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No.
60601-1:08; EN 80601-2-35; Control No.4HZ8
Classified under IEC 60601-1 Guidelines (and other
national versions of the Guidelines) as Class I, Type BF,
Ordinary equipment, Continuous operation� Not suitable
for use in the presence of flammable anesthetic mixtures
with air or with oxygen or nitrous oxide� Classified by
Underwriters Laboratories Inc� with respect to electric
shock, fire and mechanical hazards only, in accordance
with UL 60601-1, EN 80601-2-35 and in accordance with
Canadian/CSA C22�2, No� 601�1� Classified under the
Medical Device Directive as a Class IIb device�
Electrical Characteristics
blower motor
Airflow: 7-13 cfm (3�3-6�1
L
/
S
)
power consumption
Peak: 600W, Average: 340W
leakage current
Meets IEC 60601-1 and UL 60601-1 requirements
heating element
480W Resistive
power cord
15-foot, SJT, 3 cond�, 15A
4�6 m, HAR, 3 cond�, 10A
device ratings
100-120 VAC, 50/60 Hz, 4�6A
220-240 VAC, 50/60 Hz, 2�8 A
fuses
Fast Acting, 6�0A (120 VAC); 4�0A (220-230 VAC), High
Breaking Capacity
temperature controller cord
32" from hose collar, 4 cond�, Max� voltage: 5V