Bair Paws Model 875 Warming Unit Service Manual
18
Check the 3M
TM
Bair Paws
TM
system website to ensure you have the most recent version of this document.
www.bairpaws.com reorder #202036A
Replacing the Filter
Service Frequency
Every six months�
Tools/Equipment Needed
•
Medium Phillips screwdriver
•
Replacement filters (available from 3M Patient Warming Customer Service)
Method
1� Disconnect the warming unit from the power source�
2� Remove the two screws that attach the case bottom (dark purple) to the warming unit�
3� Slide the case bottom down the hose, remove the used filter, and replace with a new
filter�
4� Reposition the case bottom (with the new filter) and attach to the warming unit with
the two screws�
5� Reconnect the warming unit to the power source and turn the temperature controller
on to make sure the warming unit functions properly before returning to service�
Note: Verify calibration and test the OT circuit if it has been longer than six months since it
was last serviced�