Bair Paws Model 875 Warming Unit Service Manual
6
Check the 3M
TM
Bair Paws
TM
system website to ensure you have the most recent version of this document.
www.bairpaws.com reorder #202036A
Proper Use and Maintenance
Arizant Healthcare Inc�, a 3M company assumes no responsibility for the reliability,
performance, or safety of the unit if any of the following events occur:
•
Modifications or repairs are performed by unqualified personnel�
•
The warming unit is used in a manner other than that described in the Operator’s
Manual or Service Manual�
•
The warming unit is installed in an environment that does not meet the appropriate
electrical and grounding requirements�
Read Before Servicing Unit
All repair, calibration, and servicing of the Model 875 warming unit must be performed by
qualified, medical equipment service technicians who are familiar with good practice for
medical device repair� If the warming unit does not require the manufacturer’s attention,
3M Patient Warming will ship replacement parts to your location� Perform all repairs and
maintenance in accordance with the instructions provided with the replacement parts�
Safety Inspection
Perform a safety inspection after making repairs to the Model 875 unit and before returning
the unit to service� A safety inspection should include calibrating the operating temperature
settings and testing the over-temperature detection function as described in the Service
Procedures section as well as testing for leakage current and continuity check on safety ground�