Replacing the Fuses
Tools/Equipment Needed
•
Small Phillips screwdriver
•
Medium Phillips screwdriver
Method
1� Disconnect the warming unit from the power source�
2� Remove the two screws that attach the case bottom (dark purple) to the warming unit,
and slide the case bottom down the hose�
3� Remove the three screws on the back of the warming unit that hold the enclosures
together (see Figure 12)�
4� Flip the unit over and remove the cover� This will expose the control board�
5� Remove and replace each with the same fuse rating�
6� Replace the cover�
7� Install the three screws on the back of the unit�
8� Reposition the case bottom with the filter and attach to the warming unit with two
screws�
9� Plug the unit into a proper power source�
10� Calibrate unit and test OT circuit�
Fuses
Figure 16: Removing the fuses